12. LOAD CELL SIGNAL CONDITIONER BOARD JUMPER SETTINGS
Five ranges are jumper selectable. These also have to be selected in Instrument Setup software.
1. Use 2.5 mm (0.1") jumpers
.
2. Store spare jumpers on an unused jumper post.
13. THERMOCOUPLE SIGNAL CONDITIONER BOARD JUMPER SETTINGS
Seven thermocouple types are jumper selectable. These also have to be selected in Instrument
Setup software.
Board Revision A
1. Use 2.5 mm (0.1") jumpers.
2. Store spare jumpers on an unused jumper post.
Load Cell & Microvolt Ranges
FS Input Jumpers
±20.000 mV
±50.000 mV
±100.00 mV
±250.00 mV
±500.00 mV
e
a
b
c
d
Thermocouple Type E4 Jumper
J, K, E, N
T, R, S
none
j
Open T/C Indication E3 Jumper
Upscale
Downscale
h
i
h
i
j
E4
E3
3
4
1
2
PNP transistor
J5-X
J5-Y
c + b
e
- Sig
+ Sig